Rekonstruksi Fantom Toraks Menggunakan Nilai Densitas Hounsfield Unit sebagai Fantom Alternatif untuk Verifikasi Dosis Radiasi

Authors

  • Fiqi Diyona Rumah Sakit Universitas Andalas
  • Afdhal Muttaqin Departemen Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Andalas
  • Rhandyka Rafli Instalasi Onkologi Radiasi, Rumah Sakit Universitas Andalas

DOI:

https://doi.org/10.25077/jfu.14.6.540-545.2025

Keywords:

Phantom, density, thorax, CT Simulator

Abstract

Phantoms play an essential role in medical physics, particularly for radiation dose verification, yet the high cost of commercial standard phantoms often limits their availability. This study aimed to reconstruct a thoracic phantom as a replica of the human chest cavity using simple and low-cost materials. The reconstruction process began with evaluating the density values of the selected materials: cadaver bone was used as a substitute for human bone, styrofoam for lung tissue, and plasticine for soft tissues such as muscle and fat. Density measurements showed that cadaver bone ranged from +350 HU to +1000 HU, styrofoam from –700 HU to –950 HU, and plasticine from +39 HU to +234 HU. The skeletal-based reconstruction successfully represented the anatomical alignment of thoracic organs, as confirmed by CT simulator imaging. These findings indicate that the reconstructed thoracic phantom provides a feasible and cost-effective alternative to expensive commercial phantoms and can be effectively utilized for radiation dose verification in medical physics applications.

References

Boone, J. M., Strauss, K. J., Cody, D. D., McCollough, C. H., McNitt-Gray, M. F., & Toth, T. L. (2011). Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations (AAPM Report No. 204). College Park, MD: American Association of Physicists in Medicine.

Carrisa, F., Nurhayati, R., & Lestari, D. (2023). Pengembangan phantom radiologi menggunakan teknologi 3D printing berbahan PLA dan nilon dengan variasi infill density (Skripsi). Universitas Sebelas Maret, Surakarta.

Farizka, I. (2017). Analisis nilai densitas tulang rahang berdasarkan usia menggunakan Cone Beam Computed Tomography (CBCT) [Disertasi, Universitas Padjadjaran].

Faturrohmaniah, A., Darinafitri, I., & Nugroho, R. S. (2024). Perbedaan kepadatan mineral tulang menggunakan nilai Hounsfield Unit Corpus Lumbalis dan Collum Femur pada pasien kanker rektum sebelum dan sesudah kemoterapi kemoradiasi [Magister Thesis, Universitas Brawijaya].

Gevenois, P. A., de Maertelaer, V., De Vuyst, P., Zanen, J., & Yernault, J. C. (1995). Limits of normality of quantitative thoracic CT analysis. European Respiratory Journal, 8(5), 843–852. https://doi.org/10.1183/09031936.95.08050843

Hosseini, S. A., Khorshidi, A., & Shafiei, S. (2020). Development of a patient-specific 3D-printed thorax phantom based on CT data for imaging validation. Physica Medica, 80, 215–223. https://doi.org/10.1016/j.ejmp.2020.11.006

Globocan, 2020, Global Burden of Cancer Study: Cancer in Indonesia, Global Cancer Observatory.

KEMENKES, 2018, Pedoman Nasional Pelayanan Kedokteran (PNPK) Tata Laksana Kanker Payudara, In DJPK Kementerian Kesehatan Republik Indonesia (Vol. 6, Issue 1).

Kwon, J. Y., Han, J. H., & Kim, S. H. (2025). Development of a 3D-printed chest phantom with pulmonary nodules for low-dose CT optimization. Applied Sciences, 15(1), 309. https://doi.org/10.3390/app15010309

Mizutani, J., Suzuki, T., Matsumoto, H., Yamashita, K., & Hasegawa, T. (2020). Cortical bone trajectory and traditional trajectory: A radiological evaluation of screw fixation in the lumbar spine. Journal of Orthopaedic Surgery and Research, 15(1), 302. https://doi.org/10.1186/s13018-020-01821-0

Radiopaedia.org. (2023). Hounsfield unit – Radiology Reference Article. Retrieved August 27, 2025, from https://radiopaedia.org/articles/hounsfield-unit

Radiopaedia.org. (2023). Hounsfield unit – Radiology Reference Article. Retrieved August 27, 2025, from https://radiopaedia.org/articles/hounsfield-unit

Schmidt, B., Flohr, T. G., Heismann, B., Raupach, R., Fuchs, T., Nielsen, T., & Padole, A. (2024). Comparison of Hounsfield Unit variability between two photon-counting CT systems: A phantom study applied to lung CT. Physica Medica, 118, 103845. https://doi.org/10.1016/j.ejmp.2024.103845

Smith, B. D., Bentzen, S. M., Correa, C. R., dkk., (2011). Fractionation for whole breast irradiation: An American society for radiation oncology (ASTRO) evidence-based guideline. International Journal of Radiation Oncology Biology Physics, 81(1), 59–68. https://doi.org/10.1016/j.ijrobp.2010.04.042

Wang, M., Chen, Y., Zhang, H., Li, J., & Xu, X. (2024). Bone density of the cervical, thoracic and lumbar spine measured using Hounsfield units of computed tomography: Results of 4350 vertebras. BMC Musculoskeletal Disorders, 25(1), 178. https://doi.org/10.1186/s12891-024-07324-1

Wikipedia. (2024). Hounsfield scale. Retrieved August 27, 2025, from https://en.wikipedia.org/wiki/Hounsfield_scale

Xue, C., Sun, G., Wang, N., Liu, X., He, G., Wei, Y., & Xi, Z. (2024). Value of Hounsfield units measured by chest computed tomography for assessing bone density in the thoracolumbar segment of the thoracic spine. Asian Spine Journal, 18(3), 336–345. https://doi.org/10.31616/asj.2023.0438

Yunus, B., & Murtala, B. (2010). Pemanfaatan Hounsfield Unit pada CT-scan dalam menentukan kepadatan tulang rahang untuk pemasangan implan gigi. Journal of Dentomaxillofacial Science, 9(1). https://doi.org/10.15562/jdmfs.v9i1.230

Zhou, W., Li, Y., Wang, Y., & Sun, Z. (2023). CT and MR for bone mineral density and trabecular bone score assessment in osteoporosis evaluation. Quantitative Imaging in Medicine and Surgery, 13(10), 6361–6372. https://doi.org/10.21037/qims-23-389

Downloads

Published

12-11-2025

How to Cite

Diyona, F., Muttaqin, A., & Rafli, R. (2025). Rekonstruksi Fantom Toraks Menggunakan Nilai Densitas Hounsfield Unit sebagai Fantom Alternatif untuk Verifikasi Dosis Radiasi. Jurnal Fisika Unand, 14(6), 540–545. https://doi.org/10.25077/jfu.14.6.540-545.2025

Issue

Section

Articles