Pengembangan Quality Control Phantom Berbahan Akrilik dan Aluminium
DOI:
https://doi.org/10.25077/jfu.15.5.470-475.2026Keywords:
Phantom, Quality Control, Grey Level, X-rayAbstract
The high cost of phantoms for diagnostic image quality control presents a challenge, especially for hospitals in remote areas. This study developed a low-cost, simplified phantom using acrylic and aluminum materials. Circular acrylic sheets, each 3 mm thick, were stacked into 10 layers and drilled with 15 holes, each filled with aluminum disks of varying thickness from 1 mm to 15 mm. The phantom was exposed to X-rays using tube voltages of 50 kV, 60 kV, and 70 kV for 5 ms at a tube current of 160 mA. The resulting radiographic images were analyzed using ImageJ to assess grayscale values. Results showed that images obtained at 60 kV provided the best contrast, with clear differentiation of aluminum thicknesses and separation from the background. At 50 kV, aluminum layers thicker than 10 mm were not easily distinguishable due to insufficient X-ray energy, while at 70 kV, thinner layers (1–5 mm) lacked visibility due to over-penetration. These findings suggest that the developed phantom is adequately responsive to exposure parameter changes and has potential for use in basic radiographic imaging evaluations and educational settings.
References
Akhlaghi, P., Miri Hakimabad, H., & Rafat Motavalli, L. (2015). Determination of tissue equivalent materials of a physical 8-year-old phantom for use in computed tomography. Radiation Physics and Chemistry, 112, 169–176. https://doi.org/10.1016/j.radphyschem.2015.03.030
Alves, A. F. F., Miranda, J. R. de A., Bacchim Neto, F. A., Duarte, S. B., & Pina, D. R. de. (2015). Construction of pediatric homogeneous phantoms for optimization of chest and skull radiographs. European Journal of Radiology, 84(8), 1579–1585. https://doi.org/10.1016/j.ejrad.2015.05.015
Bor, D., Unal, E., & Uslu, A. (2015). Comparison of different phantoms used in digital diagnostic imaging. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 795, 160–166. https://doi.org/10.1016/j.nima.2015.05.013
Bustos Flores, M., Prata Mourão, A., & Chevalier del Rio, M. (2022). Breast phantom made of acrylic slabs for tests in mammography DR. Applied Radiation and Isotopes, 188, 110389. https://doi.org/10.1016/j.apradiso.2022.110389
Danz, J. C., Flueck, H. P., Campus, G., & Wolf, T. G. (2023). Computed vs. film-based radiographs’ contour artifacts influence diagnosis of secondary caries. European Journal of Radiology, 166, 111004. https://doi.org/10.1016/j.ejrad.2023.111004
Gumrukcu, Z., Bayrak, S., & Balaban, E. (2025). Clinical and radiographical evaluation of the effects of two different flap designs in third molar surgery: A single blind randomized clinical trial. Current Problems in Surgery, 62, 101675. https://doi.org/10.1016/j.cpsurg.2024.101675
Mehdizadeh, M., Tavakoli, M. B., & Moghaddam, F. G. (2020). Investigating the Effect of Different Metal Sheets on Reducing Radiation Behind Intraoral Digital Phosphor Plates During Intraoral Radiography using Gray Value. The Open Dentistry Journal, 14(1), 704–710. https://doi.org/10.2174/1874210602014010704
Mohammed Ali, A., & Al-Murshedi, S. (2023). Low-cost chest paediatric phantom for dose optimisation: construction and validation. Radiología (English Edition), 65(4), 327–337. https://doi.org/10.1016/j.rxeng.2022.11.001
Mohammed Ali, A., Hogg, P., Johansen, S., & England, A. (2018). Construction and validation of a low cost paediatric pelvis phantom. European Journal of Radiology, 108, 84–91. https://doi.org/10.1016/j.ejrad.2018.09.015
Nguyen, V.-H., Tran, N.-L., Vu, P.-T., Le, T.-L., & Pham, N.-H. (2016). Development of digital image processing software for Vikomed X-ray equipment. 2016 International Conference on Biomedical Engineering (BME-HUST), 41–44. https://doi.org/10.1109/BME-HUST.2016.7782099
Souza, S. A. S. de, Alves, A. F. F., Mamprim, M. J., & Pina, D. R. (2022). Quality and dose optimization in canine chest radiography using a digital radiography system. Radiation Physics and Chemistry, 195, 110085. https://doi.org/10.1016/j.radphyschem.2022.110085
Sriwahyuni. (2017). PENGARUH TEGANGAN TABUNG (KV) TERHADAP KUALITAS CITRA RADIOGRAFI PESAWAT SINAR-X DIGITAL RADIOGRAPHY (DR) PADA PHANTOM ABDOMEN. SPEKTRA: Jurnal Fisika dan Aplikasinya, 2(2), 113–118. https://doi.org/10.21009/SPEKTRA.022.04
Velonis, M., Papanastasiou, E., Hatziioannou, K., Siountas, A., Kamperis, E., Papavasileiou, P., Koukourakis, M. I., & Seimenis, I. (2023). Dose optimization of 2D X-ray image acquisition protocols in image-guided radiotherapy. Physica Medica, 115, 103161. https://doi.org/10.1016/j.ejmp.2023.103161
Yalcin, A., & Olgar, T. (2024). Developing a novel phantom for image quality evaluation in digital radiography. Radiation Physics and Chemistry, 222, 111868. https://doi.org/10.1016/j.radphyschem.2024.111868
Yen, J.-Y., Hsu, H.-J., Lai, Y.-L., Chou, I.-C., Chen, Y.-C., & Lee, S.-Y. (2024). Efficacy of customized crown-level position jig in measuring peri-implant crestal bone level on periapical radiographs: An in vitro study. Journal of Dental Sciences, 19(1), 338–344. https://doi.org/10.1016/j.jds.2023.06.021
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Alfieta Rohmaful Aeni, Danang Noeramarullah Yudhistira

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.