Analisis Pengaruh Penggunaan Physical Wedge Filter pada Dose Profile dan Percentage Depth Dose Pesawat Radioterapi Clinax-Cx di Rumah Sakit Universitas Andalas
DOI:
https://doi.org/10.25077/jfu.14.2.184-190.2025Keywords:
dose profile, LINAC, Percent Depth Dose, wedge filterAbstract
A study has been conducted on the effect of using a physical wedge filter on the dose profile and Percentage Depth Dose (PDD) of the CLINAC-CX teletherapy device at Andalas University Hospital. This study aims to determine the shape of the dose profile and PDD curves formed from the use of physical wedges. Measurements were made with variations in photon beam energy (6 MV and 10 MV) and variations in the angle of the physical wedge filter (15°, 30°, 45°, and 60°). Data processing begins by calculating incomplete data from the LINAC using interpolation techniques. The data is then displayed in the form of a curve or graph using Microsoft Excel. The results of the PDD curve measurement showed no significant difference in the curve formed using or without a wedge. The resulting zmax did not differ much from each other, where for a 6 MV beam without a wedge it was 13.8 mm while the average wedge variation was 14.72 mm. The 10 MV beam variation showed a value of 23.4 mm and an average of 24 mm. The dose profile formed is influenced by the use of wedge, the comparison of the smallest and largest symmetry, flatness, and penumbra values. The smallest values at 6 MV irradiation without wedge are respectively 2.1%, 3.1%, 7.1 mm, and 7.3 mm, while the largest with a wedge angle of 60°, 32.4%, 58.3%, 15.8 mm. The wedge factor between 6 MV and 10 MV shows no difference.
References
Ahmad, M., Hussain, A., Muhammad, W., Rizvi, S. Q. A., & Matiullah. (2010). Studying wedge factors and beam profiles for physical and enhanced dynamic wedges. Journal of Medical Physics, 35(1), 33–41. https://doi.org/10.4103/0971-6203.57116.
Akasaka, H., Mukumoto, N., Nakayama, M., Wang, T., Yada, R., Shimizu, Y., Osuga, S., Wakahara, Y., & Sasaki, R. (2017). A Comparison of Physical vs. Nonphysical Wedge Modalities in Radiotherapy. InTech.
Chakrabarti, J., & Mitra, S. (2017). Cancer and Noncoding RNAs (1st ed.). Elsevier Science.
Geraily, G., Mirzapour, M., Mahdavi, S. R., & Allahverdi, M. (2014). Monte Carlo study on beam hardening effect of physical wedges. 12(3).
Haffty, B. ., & Wilson, L. . (2009). Handbook of Radiation Oncology: Basic Principles and Clinical Protocols. Jones & Bartlett Publisher.
Khan, F. M., & Gibbons, J. P. (2014). Khan’s the Physics of Radiation Therapy. Lippincott Williams & Wilkins.
Milvita, D., Mahyudin, A., & Alvionita, V. (2018). Komunikasi Fisika Indonesia Analisis Nilai Percentage Depth Dose ( Pdd ) Terhadap Variasi Kedalaman Target Dan Luas Lapangan Penyinaran Menggunakan Pesawat Linac-Cx. 15(02).
Perez, C. ., Brady, L. ., & Halperin, E. (2008). Principles and practice of radiation oncology (Fifth). Lippincott Williams & Wilkins.
Shamsi, A., Birgani, M. J. T., Behrooz, M. A., Arvandi, S., Fatahiasl, J., Maskny, R., & Abdalvand, N. (2016). Determination of an effective wedge angle by combination of two arbitrary universal wedge fields in radiation therapy of cancer patients with megavoltage photon beams. Asian Pacific Journal of Cancer Prevention, 17(1), 197–200. https://doi.org/10.7314/APJCP.2016.17.1.197
Zabihzadeh, M., Fadaei, M., Hoseini, S. M., Arvandi, S., & Tahmasbi, M. (2020). Characterization of wedge factors and dose distributions in radiotherapy with symmetric and asymmetric physical wedged beams of 6 MV photon beam. Iranian Journal of Medical Physics, 17(3), 213–219. https://doi.org/10.22038/ijmp.2019.41275.1595