Perbandingan Hasil Simulasi Software Primo dengan Pengukuran Eksperimen pada Quality Assurance Berkas Foton Linear Accelerator Varian Clinac CX

Authors

DOI:

https://doi.org/10.25077/jfu.15.4.430-436.2026

Keywords:

Analisa Percentage Depth Dose (PDD), Dose profile, Gamma, Linear Accelerator (Linac), quality assurance

Abstract

This study aims to compare the results of Monte Carlo simulations using PRIMO software with the results of experimental measurements in the QA process of a 6 MV Varian Clinac CX Linear Accelerator (LINAC) photon beam based on the parameters of Percentage Depth Dose (PDD), dose profile, and gamma index. Dose distribution measurements were performed using a water phantom at a Source to Surface Distance (SSD) of 100 cm with field sizes of 4×4 cm², 10×10 cm², and 40×40 cm². Monte Carlo simulations were performed using PRIMO software by modeling the geometry of the Varian Clinac CX LINAC head and water phantom according to the experimental conditions. The simulation and measurement results were analyzed through PDD curves, dose profiles, and quantitative evaluation using the gamma index with a tolerance criterion of 2%/2 mm. The results showed that the simulated PDD curves were in good agreement with the measurement results. The gamma passing rate values obtained were 97.98% for the 4×4 cm² field, 97.43% for the 10×10 cm² field, and 87% for the 40×40 cm² field. The decrease in conformity in large fields is influenced by an increase in the contribution of lateral scattering. The dose profile analysis results show conformity in the central area of the field, while greater differences occur in the penumbra and shallow depths due to steep dose gradients and detector resolution limitations. PRIMO-based Monte Carlo simulation can be relied upon as an independent verification instrument in the routine quality assurance of 6 MV photon beams from the Varian Clinac CX LINAC

References

Altuwayrish, A., Ghorbani, M., Bakhshandeh, M., Roozmand, Z., HoseiniGhahfarokhi, M., 2022, Comparison of PRIMO Monte Carlo code and Eclipse treatment planning system in calculation of dosimetric parameters in brain cancer radiotherapy, Reports of Practical Oncology and Radiotherapy, Vol. 27, Hal. 863–874, DOI: 10.5603/rpOr.a2022.0091.

American Cancer Society, 2024, 2024-Cancer-Facts-and-Figures-Acs, Cancer Facts and Figures.

Brualla, L., Rodriquez, M., Sempau, J., 2020, PRIMO User Manual, Germany.

Daniel A. Low, a) William B. Harms, Sasa Mutic, and J.A.P., 1998, Prescribing, Recording, and Reporting Photon Beam Therapy: Supplement to ICRU Report 50, ICRU Press, Besthesda, Maryland, USA.

Darafsheh, A., 2021. Radiation therapy dosimetry: A practical handbook.CRC Press. https://doi.org/10.1201/9781003037248

Efendi, M.A., Funsian, A., Chittrakarn, T., Bhongsuwan, T., 2017, Monte carlo simulation of 6 MV flattening filter free photon beam of truebeam STx LINAC at songklanagarind hospital, Sains Malaysiana, Vol. 46, Hal. 1407– 1411, DOI: 10.17576/jsm-2017-4609-08.

International Commission on Radiation Units and Measurements (ICRU). (1993). Prescribing, recording, and reporting photon beam therapy .

Klein, E.E., Hanley, J., Bayouth, J., Yin, F.F., Simon, W., Dresser, S., Serago, C., Aguirre, F., Ma, L., Arjomandy, B., Liu, C., Sandin, C., Holmes, T., 2009, Task group 142 report: Quality assurance of medical acceleratorsa, Medical Physics, Vol. 36, Hal. 4197–4212, DOI: 10.1118/1.3190392.

Mayles, P., Nahum, A. E., & Rosenwald, J.-C., 2007. Handbook of radiotherapy physics: Theory and practice. CRC Press.

Podgorsak, E.B., n.d., Radiation Oncology Physics : A Handbook for Teachers and Students.

Rodriguez, M., Sempau, J., Fogliata, A., Cozzi, L., Sauerwein, W., Brualla, L., 2015, A geometrical model for the Monte Carlo simulation of the TrueBeam linac, Physics in Medicine and Biology, Vol. 60, Hal. N219–N229, DOI: 10.1088/0031-9155/60/11/N219.

Sarin, B., Bindhu, B., Saju, B., Nair, R. K., 2020, Validation of PRIMO Monte Carlo model of Clinac iX 6 MV photon beam, Journal of Medical Physics, Vol. 45(1), Hal. 24–35, DOI: 10.4103/jmp.JMP_75_19

Sempau, B.R., 2015, User ’ s Manual.

Susworo, R. (2007). Radioterapi: Dasar-dasar radioterapi, tata laksana radioterapi penyakit kanker (Edisi ke-5). UI-Press.

Stephen Morris, Tom Roques, shahreen Ahmad, S.L., 2023, No Title, Edisi fifth Edit, 2 Oktober 2023, Boca Raton, Florida, USA.

Thomadsen, B. R., Rivard, M. J., & Butler, W. M., 2015. Brachytherapy physics (2nd ed.). Medical Physics Publishing

Williams, J. R., & Thwaites, D. I., 1993. Radiotherapy physics in practice. Oxford University Press.

Downloads

Published

06-08-2026

How to Cite

Roelyant, W., Adrial, R., & Efendi, M. A. (2026). Perbandingan Hasil Simulasi Software Primo dengan Pengukuran Eksperimen pada Quality Assurance Berkas Foton Linear Accelerator Varian Clinac CX. Jurnal Fisika Unand, 15(4), 430–436. https://doi.org/10.25077/jfu.15.4.430-436.2026

Issue

Section

Articles