Estimasi Laju Dosis Radiasi yang Lolos Melalui Dinding Primer dan Sekunder Berdasarkan Beban Kerja Pesawat Teleterapi Co-60 di Instalasi Radioterapi RSUP. Dr. M. Djamil Padang (Studi Kasus Januari - Juni 2019)

Authors

  • Santri Ramadhani Jurusan Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Andalas
  • Dian Milvita Jurusan Fisika, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Andalas
  • Heru Prasetio PTKMR-BATAN
  • Mursiyatun Mursiyatun PTKMR-BATAN

DOI:

https://doi.org/10.25077/jfu.9.3.284-291.2020

Abstract

Telah dilakukan penelitian tentang estimasi laju dosis radiasi yang lolos melalui dinding primer dan sekunder, berdasarkan beban kerja pesawat teletrapi Co-60 di instalasi radioterapi RSUP. Dr. M. Djamil Padang. Penelitian menggunakan data sekunder berupa data dosis pasien, pekerja radiasi periode Januari-Juni 2019 dan data desain dinding ruang penyinaran Co-60. Metode perhitungan menggunakan persamaan analitik dari SRS No. 47 IAEA. Hasil yang didapatkan adalah nilai beban kerja pesawat Co-60 per minggu berkisar antara (6,4-288) Gy m2 per minggu. Nilai ini masih berada di bawah rekomendasi NCRP No. 49. Hasil perhitungan estimasi laju dosis radiasi yang lolos melalui dinding primer berkisar antara (5,35 × 10-9–1,99 × 10-6) Sv per minggu, sedangkan estimasi laju dosis radiasi yang lolos melalui dinding sekunder akibat radisi bocor berkisar antara (2,27 × 10-14- 2,41 × 10-6) Sv per minggu dan melalui dinding sekunder akibat radiasi hambur berkisar antara (1,56 × 10-17- 1,94 × 10-9) Sv per minggu. Nilai estimasi laju dosis radiasi yang didapatkan masih berada di bawah nilai batas dosis (NBD) yang ditetapkan oleh BAPETEN No. 3 Tahun 2013 pasal 30 dan 31 tentang limitasi dosis radiasi untuk pekerja radiasi dan masyarakat umum.

 

Research has been conducted on the estimation of the rate of radiation doses that pass through the primary and secondary walls, based on the workload of the Co-60 teletherapy instrument in the radiotherapy installation of RSUP. Dr. M. Djamil Padang. The research using secondary data, that are the patient dose and radiation workers for the January-June 2019 period and Co-60 wall design data. The method of calculation was analytic equations from Safety Report Series (SRS) International Atomic Energy Agency (IAEA) No. 47. The result show that the workload value of Co-60 ranges from (6.4-288) Gy m2 week. This value was still lower than the recommendation of NCRP No. 49. The results of the calculation of the estimated radiation dose rate that passes through the primary wall obtained ranged between (5.35 × 10-9-1.99 × 10-6) Sv per week, while through the secondary wall due to the leakage of radiation ranged between (2.27× 10-14–2.41 × 10-6) Sv per week and through the secondary wall due to scattering radiation ranged between (1.56 × 10-17–1.94 × 10-9) Sv per week. The estimated value of the radiation dose obtained is still lower than the value of dose limit (NBD) determined by BAPETEN No. 3 of 2013 articles 30 and 31 concerning the limitation of radiation doses for radiation workers and general public.

References

Arifin, S., Milvita, D., Nuraeni, N., dan Yuliati, H., Pemetaan Laju Dosis Radiasi Menggunakan TLD-100 dengan Sumber Radiasi Cobalt-60 di Instalasi Radioterapi RSUP. Dr. M. Djamil Padang, Jurnal Fisika Unand, 4(3), hal.208-211, 2015.

BAPETEN, Peraturan Kepala BAPETEN No.3 Tahun 2013 tentang Keselamtan Radiasi dalam Penggunaan Radioterapi, (Jakarta, 2013)

BAPETEN, Peraturan Kepala BAPETEN No.4 Tahun 2013 tentang Proteksi dan Keselamatan Radiasi dalam Pemanfaatan Tenaga Nuklir, (Jakarta, 2013).

Cember, H., Introduction to Health Physics, McGraw-Hill Publishing Company Ltd., (New Delhi, 1983)

IAEA, Radiation Protection in The Design of Radiotherapy Facilities, Safety Report Series No. 47, (Vienna, 2006).

Kristiyanti, Santoso, B., Jalil, A., dan Sukandar, Perancang Ruangan Radioterapi Eksternal Menggunakan Co-60, Jurnal Perangkat Nuklir, 06(02), hal.95-101, 2012.

NCRP, Structural Shielding Design and Medical Use of X Rays and Gamma Rays of Energies Up To 10 MeV, NCRP No. 49, (Washington, 1976).

NCRP, Structural Shielding Design And Evaluation for Megavlotage X and Gamma Ray Radiotherapy Facilities, NCRP No. 151, (Maryland, 2005).

Downloads

Published

2020-08-14

How to Cite

Ramadhani, S., Milvita, D., Prasetio, H., & Mursiyatun, M. (2020). Estimasi Laju Dosis Radiasi yang Lolos Melalui Dinding Primer dan Sekunder Berdasarkan Beban Kerja Pesawat Teleterapi Co-60 di Instalasi Radioterapi RSUP. Dr. M. Djamil Padang (Studi Kasus Januari - Juni 2019). Jurnal Fisika Unand, 9(3), 284–291. https://doi.org/10.25077/jfu.9.3.284-291.2020